RIGOL Chapter 2 Command System
2-148 DSA800E Programming Guide
[:SENSe]:TPOWer:AVERage:TCONtrol
Syntax
[:SENSe]:TPOWer:AVERage:TCONtrol EXPonential|REPeat
[:SENSe]:TPOWer:AVERage:TCONtrol?
Description
Set the average mode of the T-power measurement.
Query the average mode of the T-power measurement.
Parameter
Name
Type
Range
Default
--
Keyword
EXPonential|REPeat
EXPonential
Explanation
EXPonential: exponential average
REPeat: repeat average
When exponential average is selected, the result is the exponential average of the latest N (specified by the
[:SENSe]:TPOWer:AVERage:COUNt
command) measurement results.
When repeat average is selected, the result is the arithmetic average of the latest N (specified by the
[:SENSe]:TPOWer:AVERage:COUNt
command) measurement results.
This command is only available when the T-power measurement is enabled.
Return Format
The query returns EXP or REP.
Example
The command below sets the average mode to repeat average.
:SENSe:TPOWer:AVERage:TCONtrol REPeat
The query below returns REP.
:SENSe:TPOWer:AVERage:TCONtrol?
[:SENSe]:TPOWer:LLIMit
Syntax
[:SENSe]:TPOWer:LLIMit <time>
[:SENSe]:TPOWer:LLIMit?
Description
Set the start line of the T-power measurement.
Query the start line of the T-power measurement.
Parameter
Name
Type
Range
Default
<time>
Consecutive Real Number
0 us to the stop line
0 us
Explanation
This command is only available when the T-power measurement is enabled.
Return Format
The query returns the start line in scientific notation and the unit is s.