EasyManuals Logo

Rigol DSA832E Programming Guide

Rigol DSA832E
243 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #153 background imageLoading...
Page #153 background image
Chapter 2 Command System RIGOL
DSA800E Programming Guide 2-131
[:SENSe]:FREQuency:STOP
Syntax
[:SENSe]:FREQuency:STOP <freq>
[:SENSe]:FREQuency:STOP?
Description
Set the stop frequency.
Query the stop frequency.
Parameter
Name
Type
Range
Default
<freq>
Consecutive Real Number
0 Hz to 3.2 GHz
3.2 GHz
Return Format
The query returns the stop frequency in integer and the unit is Hz.
Example
The command below sets the stop frequency to 10 MHz.
:SENSe:FREQuency:STOP 10000000
The query below returns 10000000.
:SENSe:FREQuency:STOP?
[:SENSe]:HDISt:AVERage:COUNt
Syntax
[:SENSe]:HDISt:AVERage:COUNt <integer>
[:SENSe]:HDISt:AVERage:COUNt?
Description
Set the number of averages of the harmonic distortion measurement.
Query the number of averages of the harmonic distortion measurement.
Parameter
Name
Type
Range
Default
<integer>
Integer
1 to 1000
10
Explanation
This command is only available when the harmonic distortion measurement is enabled.
Return Format
The query returns the number of averages in integer.
Example
The command below sets the number of averages to 100.
:SENSe:HDISt:AVERage:COUNt 100
The query below returns 100.
:SENSe:HDISt:AVERage:COUNt?

Table of Contents

Other manuals for Rigol DSA832E

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DSA832E and is the answer not in the manual?

Rigol DSA832E Specifications

General IconGeneral
BrandRigol
ModelDSA832E
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals