RIGOL Chapter 2 Command System
2-140 DSA800E Programming Guide
[:SENSe]:POWer[:RF]:ATTenuation:AUTO
Syntax
[:SENSe]:POWer[:RF]:ATTenuation:AUTO OFF|ON|0|1
[:SENSe]:POWer[:RF]:ATTenuation:AUTO?
Description
Enable or disable the auto setting mode of the input attenuation.
Query the status of the auto setting mode of the input attenuation.
Parameter
Name
Type
Range
Default
--
Bool
OFF|ON|0|1
ON|1
Return Format
The query returns 0 or 1.
Example
The command below disables the auto setting mode of the input attenuation.
:SENSe:POWer:RF:ATTenuation:AUTO OFF or :SENSe:POWer:RF:ATTenuation:AUTO 0
The query below returns 0.
:SENSe:POWer:RF:ATTenuation:AUTO?
[:SENSe]:POWer[:RF]:GAIN[:STATe]
Syntax
[:SENSe]:POWer[:RF]:GAIN[:STATe] OFF|ON|0|1
[:SENSe]:POWer[:RF]:GAIN[:STATe]?
Description
Enable or disable the preamplifier.
Query the status of the preamplifier.
Parameter
Name
Type
Range
Default
--
Bool
OFF|ON|0|1
OFF|0
Return Format
The query returns 0 or 1.
Example
The command below enables the preamplifier.
:SENSe:POWer:RF:GAIN:STATe ON or :SENSe:POWer:RF:GAIN:STATe 1
The query below returns 1.
:SENSe:POWer:RF:GAIN:STATe?

Table of Contents

Other manuals for Rigol DSA832E

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DSA832E and is the answer not in the manual?

Rigol DSA832E Specifications

General IconGeneral
BrandRigol
ModelDSA832E
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals