Chapter 2 Command System  RIGOL 
DSA800E Programming Guide  2-125 
[:SENSe]:EBWidth:MAXHold:STATe 
 
Syntax 
[:SENSe]:EBWidth:MAXHold:STATe OFF|ON|0|1  
[:SENSe]:EBWidth:MAXHold:STATe? 
 
Description
 
Enable or disable the Max Hold. 
Query the status of the Max Hold.
 
 
Parameter
 
 
Explanation 
This command is only available when the emission bandwidth measurement is enabled.  
When Max Hold is enabled, each measurement result is compared with the previous result and the 
maximum is displayed.
 
When Max Hold is disabled, the current measurement result is displayed. 
Max Hold and the average measurement function are mutually exclusive and the average measurement 
function will be automatically disabled when Max Hold is enabled.
 
 
Return Format
 
The query returns 0 or 1. 
 
Example
 
The command below enables the Max Hold. 
:SENSe:EBWidth:MAXHold:STATe ON or :SENSe:EBWidth:MAXHold:STATe 1 
 
The query below returns 1. 
:SENSe:EBWidth:MAXHold:STATe? 
 
 
 
[:SENSe]:EBWidth:XDB 
 
Syntax 
[:SENSe]:EBWidth:XDB <real_amp>  
[:SENSe]:EBWidth:XDB? 
 
Description
 
Set the value of X dB of the EBW measurement.  
Query the value of X dB of the EBW measurement.
 
 
Parameter
 
 
Explanation
 
This command is only available when the emission bandwidth measurement is enabled.  
 
Return Format
 
The query retunes the value of X dB in scientific notation.