EasyManuals Logo

Rigol DSA832E Programming Guide

Rigol DSA832E
243 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #166 background imageLoading...
Page #166 background image
RIGOL Chapter 2 Command System
2-144 DSA800E Programming Guide
[:SENSe]:SWEep:TIME:AUTO:RULes
Syntax
[:SENSe]:SWEep:TIME:AUTO:RULes NORMal|ACCuracy
[:SENSe]:SWEep:TIME:AUTO:RULes?
Description
Set the setting method of the auto sweep time.
Query the setting method of the auto sweep time.
Parameter
Name
Type
Range
Default
--
Keyword
NORMal|ACCuracy
NORMal
Return Format
The query returns NORM or ACC.
Example
The command below sets the setting method of the auto sweep time to accuracy.
:SENSe:SWEep:TIME:AUTO:RULes ACCuracy
The query below returns ACC.
:SENSe:SWEep:TIME:AUTO:RULes?
[:SENSe]:TOI:AVERage:COUNt
Syntax
[:SENSe]:TOI:AVERage:COUNt <integer>
[:SENSe]:TOI:AVERage:COUNt?
Description
Set the number of averages of the TOI measurement.
Query the number of averages of the TOI measurement.
Parameter
Name
Type
Range
Default
<integer>
Integer
1 to 1000
10
Explanation
This command is only available when the TOI measurement is enabled.
Return Format
The query returns the number of averages in integer.
Example
The command below sets the number of averages to 100.
:SENSe:TOI:AVERage:COUNt 100
The query below returns 100.
:SENSe:TOI:AVERage:COUNt?

Table of Contents

Other manuals for Rigol DSA832E

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DSA832E and is the answer not in the manual?

Rigol DSA832E Specifications

General IconGeneral
BrandRigol
ModelDSA832E
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals