RIGOL Chapter 2 Command System
2-124 DSA800E Programming Guide
[:SENSe]:EBWidth:AVERage:COUNt command) measurement results.
When repeat average is selected, the result is the arithmetic average of the latest N (specified by the
[:SENSe]:EBWidth:AVERage:COUNt command) measurement results.
This command is only available when the emission bandwidth measurement is enabled.
Return Format
The query returns EXP or REP.
Example
The command below sets the average mode to repeat average.
:SENSe:EBWidth:AVERage:TCONtrol REPeat
The query below returns REP.
:SENSe:EBWidth:AVERage:TCONtrol?
[:SENSe]:EBWidth:FREQuency:SPAN
Syntax
[:SENSe]:EBWidth:FREQuency:SPAN <freq>
[:SENSe]:EBWidth:FREQuency:SPAN?
Description
Set the span of the emission bandwidth measurement.
Query the span of the emission bandwidth measurement.
Parameter
Name
Type
Range
Default
<freq>
Consecutive Real Number
100 Hz to 3.2 GHz
2 MHz
Explanation
This command is only available when the emission bandwidth measurement is enabled.
This setting changes the span of the spectrum analyzer.
Return Format
The query returns the span in integer and the unit is Hz.
Example
The command below sets the span to 10 MHz.
:SENSe:EBWidth:FREQuency:SPAN 10000000
The query below returns 10000000.
:SENSe:EBWidth:FREQuency:SPAN?

Table of Contents

Other manuals for Rigol DSA832E

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DSA832E and is the answer not in the manual?

Rigol DSA832E Specifications

General IconGeneral
BrandRigol
ModelDSA832E
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals