185 ESS-03 Calibration GX-6000 Operator’s Manual
4 . If you were using a 5-gas (standard 4-gas plus ESS-03) calibration
cylinder to calibrate the standard 4 sensors, continue to Step 5.
If you were using a 5-gas (standard 4-gas plus PID) calibration cylinder
or a 4-gas calibration cylinder/PID-only calibration cylinder
combination to calibrate the standard 4 sensors and/or the PID sensor:
a. Remove the tubing from the rigid tube on the probe.
b. Unscrew the 5-gas (standard 4-gas plus PID), 4-gas cylinder, or PID
cylinder from the regulator.
c. Screw the ESS-03 calibration cylinder onto the demand flow
regulator. Continue with Step 6.
5 . If you were using a 5-gas (standard 4-gas plus ESS-03) calibration
cylinder to calibrate the standard 4 sensors:
a. Leave the tubing connected to the rigid tube on the probe. Continue
with Step 6.
6 . Use the AIR and SHIFT (PANIC) buttons to move the cursor next to
GAS SELECT.
AUTO
AUTO CAL
CONCENTRATION
GAS SELECT
ESCAPE