EasyManuals Logo

RKI Instruments GX-6000 User Manual

RKI Instruments GX-6000
262 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #194 background imageLoading...
Page #194 background image
185 ESS-03 Calibration GX-6000 Operator’s Manual
4 . If you were using a 5-gas (standard 4-gas plus ESS-03) calibration
cylinder to calibrate the standard 4 sensors, continue to Step 5.
If you were using a 5-gas (standard 4-gas plus PID) calibration cylinder
or a 4-gas calibration cylinder/PID-only calibration cylinder
combination to calibrate the standard 4 sensors and/or the PID sensor:
a. Remove the tubing from the rigid tube on the probe.
b. Unscrew the 5-gas (standard 4-gas plus PID), 4-gas cylinder, or PID
cylinder from the regulator.
c. Screw the ESS-03 calibration cylinder onto the demand flow
regulator. Continue with Step 6.
5 . If you were using a 5-gas (standard 4-gas plus ESS-03) calibration
cylinder to calibrate the standard 4 sensors:
a. Leave the tubing connected to the rigid tube on the probe. Continue
with Step 6.
6 . Use the AIR and SHIFT (PANIC) buttons to move the cursor next to
GAS SELECT.
AUTO
AUTO CAL
CONCENTRATION
GAS SELECT
ESCAPE

Table of Contents

Other manuals for RKI Instruments GX-6000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RKI Instruments GX-6000 and is the answer not in the manual?

RKI Instruments GX-6000 Specifications

General IconGeneral
BrandRKI Instruments
ModelGX-6000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals