21 Start Up GX-6000 Operator’s Manual
4 . If CAL REMINDER is set to ON, the screen that appears next depends
on how CAL EXPRD is set in Maintenance Mode (see “Updating the
Calibration Expired Setting” on page 150).
If the unit is due for calibration and CAL EXPRD is set to
CONFIRM TO USE, then the following screen displays and the
buzzer sounds in a double pulsing pattern.
To perform a calibration, press and release the POWER/ENTER
button. The GX-6000 will enter Calibration Mode and the LCD will
show the Calibration Mode main menu. See “Chapter 4: Calibration
Mode” on page 68 for instructions to calibrate the GX-6000. When
you are done with the calibration and exit Calibration Mode, the unit
will begin the startup sequence. If the calibration was successful, the
screen above will not appear again until the unit is due for
calibration. If the calibration was not successful, the screen above
will again appear in the startup sequence.
To continue without performing a calibration, press and release the
RESET button.
If the unit is due for calibration and CAL EXPRD is set to
CANNOT USE, then the following screen displays and the buzzer
sounds in a double pulsing pattern.
4:07
CAL DATE PAST
CONFIRM TO USE
CAL MODE : ENTER
NO : RESET
4:07
CAL DATE PAST
CANNOT USE
CAL MODE : ENTER

Table of Contents

Other manuals for RKI Instruments GX-6000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RKI Instruments GX-6000 and is the answer not in the manual?

RKI Instruments GX-6000 Specifications

General IconGeneral
BrandRKI Instruments
ModelGX-6000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals