GX-6000 Operator’s Manual Start Up • 24
• If the unit is due for bump testing and BUMP EXPRD is set to
CANNOT USE, then the following screen displays and the buzzer
sounds in a double pulsing pattern.
The GX-6000 cannot be used until a successful bump test has been
performed. Press and release the POWER/ENTER button to enter
Calibration Mode. See “Chapter 4: Calibration Mode” on page 68 for
instructions to bump test the GX-6000.
NOTE: In this situation, even if the User password function has been turned
on, no password is required to perform a bump test.
NOTE: If BUMP DISP is set to OFF, the bump test menu item will not
appear in Calibration Mode even though the instrument is
prompting you to perform a bump test. A bump test can always be
performed in Maintenance Mode, if necessary.
When you are done with the bump test and exit Calibration Mode,
the unit will begin the startup sequence. If the bump test was
successful, the screen above will not appear again until the unit is due
for bump testing. If the bump test was not successful, the screen
above will again appear in the startup sequence.
4:07
BUMP DATE PAST
CANNOT USE
CAL MODE : ENTER