EasyManuals Logo

RKI Instruments GX-6000 User Manual

RKI Instruments GX-6000
262 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #32 background imageLoading...
Page #32 background image
23 Start Up GX-6000 Operator’s Manual
5 . If BUMP REMINDER is set to ON (factory setting is OFF), the screen
that appears next depends on how BUMP EXPRD is set in Maintenance
Mode (see “Updating the Bump Expired Setting” on page 155).
If the unit is due for bump testing and BUMP EXPRD is set to
CONFIRM TO USE, then the following screen displays and the
buzzer sounds in a double pulsing pattern.
To perform a bump test, press and release the POWER/ENTER
button. The GX-6000 will enter Calibration Mode and the LCD will
show the Calibration Mode main menu. See “Chapter 4: Calibration
Mode” on page 68 for instructions to bump test the GX-6000. When
you are done with the bump test and exit Calibration Mode, the unit
will begin the startup sequence. If the bump test was successful, the
screen above will not appear again until the unit is due for bump
testing. If the bump test was not successful, the screen above will
again appear in the startup sequence.
To continue without performing a bump test, press and release the
RESET button.
NOTE: If BUMP DISP is set to OFF, the bump test menu item will not
appear in Calibration Mode even though the instrument is
prompting you to perform a bump test. A bump test can always be
performed in Maintenance Mode, if necessary.
4:07
BUMP DATE PAST
CONFIRM TO USE
CAL MODE : ENTER
NO : RESET

Table of Contents

Other manuals for RKI Instruments GX-6000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RKI Instruments GX-6000 and is the answer not in the manual?

RKI Instruments GX-6000 Specifications

General IconGeneral
BrandRKI Instruments
ModelGX-6000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals