GX-6000 Operator’s Manual Measuring Mode, Alarms 42
2 . The error code meanings are shown in the table below:
3 . If the error code is anything but 022 as shown above, the instrument
cannot be used. Contact RKI Instruments, Inc. as soon as possible.
If the error code is 022, you may press and release the RESET button to
continue into Measuring Mode if the instrument must be used
temporarily.
CAUTION: There will be no datalogging function if you operate the
instrument after a 022 system failure. Contact RKI Instruments,
Inc. as soon as possible.
Responding to a Man Down Warning 1 and Warning 2
The Man Down Warning 1 and Warning 2 alarms occur after the WARNING
1 TIME and WARNING 2 TIME, respectively, has passed since the last
movement of the instrument. See “Updating the WARNING 1 TIME Setting
on page 167 and “Updating the WARNING 2 TIME Setting” on page 167 for
instructions to change these values.
1 . Follow your established procedure for a man down warning.
2 . To silence the alarm and reset the Man Down clock, move the instrument
or press and release the RESET button.
Responding to a Man Down Alarm
The Man Down Alarm alarms occur after the ALARM TIME has passed
since the last movement of the instrument. See “Updating the ALARM TIME
Setting” on page 168 for instructions to change this value.
1 . Follow your established procedure for a man down alarm.
2 . To silence the alarm and reset the Man Down clock, press and release the
RESET button. Moving the instrument will not reset the alarm or Man
Down clock.
Table 8: Error Code Explanation
Error Code Explanation
000 ROM failure
010 RAM failure
021 FRAM failure
022 FLASH memory failure