79 • Calibrating Using the Auto Calibration Method GX-6000 Operator’s Manual
21 . If you were using a 5-gas calibration cylinder, continue to Step 22.
If you were using a 4-gas calibration cylinder:
a. Remove the tubing from the rigid tube on the probe.
b. Unscrew the 4-gas cylinder from the regulator.
c. Screw the PID calibration cylinder onto the demand flow regulator.
Continue with Step 23.
22 . If you are using a 5-gas calibration cylinder:
a. Leave the tubing connected to the rigid tube on the probe. Continue
with Step 23.
23 . Use the ▲AIR and SHIFT ▼ (PANIC) buttons to move the cursor next to
GAS SELECT.
AUTO
AUTO CAL
CONCENTRATION
GAS SELECT
ESCAPE