GX-6000 Operator’s Manual Calibrating Using the Auto Calibration Method • 80
24 . Press and release the POWER/ENTER button. The first of the Gas Select
Screens, the standard four sensors screen, will be displayed. In addition
to the standard four sensors screen, there is a separate screen for each
installed smart sensor, and an ESCAPE screen that will take you back to
the Auto Cal menu. Use the ▲AIR and SHIFT ▼ (PANIC) buttons to
scroll through the different screens.
AUTO
GAS SELECT
CH4 50 %LEL
O2 12.0 %
H2S 25.0 ppm
CO 50 ppm
AUTO
GAS SELECT
VOC 100 ppm
AUTO
GAS SELECT
ESCAPE
▲ AIR or
SHIFT ▼ (PANIC)
▲ AIR or
SHIFT ▼ (PANIC)
▲ AIR or
SHIFT ▼ (PANIC)