55 CalibrationCalibration
Perform the following procedures to register functions in [ CAL].
1. Call out the calibration menu
Press [ CAL] for three or more consecutive
seconds.
2.
Assign a calibration method to the
calibration key
[Internal Weights calibration], [Internal Weight Check]
can only be selected with W-AD/ W/ X series.
(1) Select [Calibration Key Assignment], and then
press [ OK].
(2) Select the function to be assigned, and then press
[ OK].
3. Return to the weight measurement mode
Press [ POWER]
^ [Adjustment and Tests with Internal Weights (W-AD/ W/ X Series only)] (P.46)
^ [Adjustment and Tests with External Weights] (P.47)
Using a verified balance as a legal measuring instrument in the EU:
Span calibration must be performed once the balance is installed and before using the
balance as a legal measuring instrument in the EU. The balance must be connected to power
and warmed up for at least one hour prior to span calibration and use as a legal measuring
instrument.
Using a verified balance as a legal measuring instrument in the EU:
The balance must be used within the temperature range indicated on the verification label.
CAUTION
Instructions
In case of using the Multi-stand or AP Holder, remove the Multi-stand or
AP Holder when you adjust and test your balance.