Intermittent Printing/Output (Interval Timer
Function)
The displayed weight values are automatically output at
preset intervals without having to press [ PRINT] for each
measurement. When the interval timer is set at ON,
(interval timer mark) will be displayed at the bottom of the
screen.
Cannot be Used in Combination with the Auto
Print Function
The interval timer function cannot be used in combination with the auto print function.
1. Call out the interval timer function
(1) Press [ PRINT] for three or more consecutive seconds, and then press [ OK] to
display the print setup menu.
(2) Select [Interval Timer], and then press [ OK].
(3) Select [on] or [off], and then press [ OK].
· Proceed to Step 2 if [on] has been selected.
· End operations if [off] has been selected.
Calling Out the Continuous Output Function from [MENU]
Press [ MENU], select [
System settings], press [ OK], select [Print], press [ OK],
select [Interval Timer], and then press [ OK]. This will perform the same operations.
2.
Set whether to output a value that has not been averaged as
the output value
(1) Select [Set], and then press [ OK].
(2) Select [Averaging], and then press [ OK].
(3) Select [on] or [off], and then press [ OK].
3. Set the output interval
(1) Select [Output Interval], and then press [ OK].
(2) Enter the output interval within a range of 00:00 to 99:59 minutes/seconds, and then press
[ OK]. The set value is 00:01 seconds when shipped from the factory.
When 00:00 has been Set
Continual output will be performed at an interval of approximately 100msec.
Loading weight fluctuations in the real time
Setting [Averaging] to [off]and [Output Interval] to [00:00] will enable the data to be loaded in the
real time while monitoring the volatility of the liquid.
After the interval timer has been set
Setting [Interval Timer] to [Set] will automatically set this function to valid (on).
4. Return to the weight measurement mode
Press [ POWER].
The interval timer mark changes to a
reverse display during output
*
*
*
*
*

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu AP W-AD Series and is the answer not in the manual?

Shimadzu AP W-AD Series Specifications

General IconGeneral
BrandShimadzu
ModelAP W-AD Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals