137
99
Connections and Communications with External EquipmentConnections and Communications with External Equipment
1
3
5
2
4
6
7
8
12
14
13
9
10
11
Communication settings
The Communication settings menu is used to set the communication specifications when a personal
computer or printer, etc., are connected.
The default setting is [Standard].
Four other [Standard Settings] are also available that are the most frequently used Communication
setting combinations.
Selecting one of the combinations from [Standard Setting 1] through [Standard Setting 5] enables all
of the baud rate (communication speed), parity (bit length), stop bit, handshake, and data format and
delimiter parameters to be set with a single operation.
^ [Standard Settings] (this page)
Each item can be set as required by the user.
^ [User Settings] (P.138)
[Communication setting Table]
Standard Expansion Type M Type S Type A User Settings
Supporting
Manufacturers, etc.
Shimadzu
(Standard)
Shimadzu*
(With response)
Mettler Sartorius A&D
Baud Rate
(Communication Speed)
1200 1200 2400 1200 2400 User-specified
Parity (Bit length) None (8) None (8) Even (7) Odd (7) Even (7) User-specified
Stop Bit 1 1 2 2 2 User-specified
Handshake OFF OFF OFF Hard OFF User-specified
Data Format Shimadzu
Standard
Shimadzu
Standard
Mettler
Standard
Sartorius
Standard
A&D
Standard
User-specified
Delimiter C/R C/R C/R+L/F C/R+L/F C/R+L/F User-specified
* It is possible to respond to commands received from a personal computer. OK (C/R) is returned
when received normally, and NG (C/R) is returned when an abnormality occurs.
Standard Settings
The standard settings can be selected from the pre-set Communication settings available.
1. Start Communication setting
(1) Press [ PRINT] while in the weight measurement mode for three or more consecutive
seconds.
(2) Select [Communication settings], and then press [ OK].
2. Select the Communication settings from the standard settings
(1) Select either [RS232C] or [USB] as the connector terminal for which Communication settings
are to be made, and then press [ OK].
(2) Select either [Standard], [Expanded], [M Format], [S Format], [A Format] or [User Settings],
and then press [ OK].
3. Return to the weight measurement mode
Press [ POWER].