33
1
3
5
2
4
6
7
8
12
14
13
9
10
11
Measurements (Standard measurement Mode)
1. Check the Standard measurement mode
Standard measurement Mode
A function for measuring samples without using any special functions.
Perform the following operations in accordance with the status of the balance in order to activate the
Standard measurement mode.
Balance Status Activating the Weight Measurement Mode
OFF, STAND-BY display
Press [ POWER]
Application Function Mode is activated
Press [ DOWN].
Press [ MENU], press [ LEFT], press [ UP], select [ Select
Mode], press [ OK], select [ Standard measurement], and then press
[ OK].
Menu is displayed
Press [ POWER].
Numeral input is activated
Press [ POWER] until the weight measurement mode is activated.
If [OL (OVERLOAD)] or [-OL (OVERLOAD)], etc., is displayed during measurements
Indicates that a load exceeding the weighing capacity is being measured, or that the weighing
pan is not properly in place.
2. Place the tare (container) on the pan
(1) Open the glass draft shield.
(2) Place the tare (container) on the pan, and then close the glass draft shield.
3.
When the display stabilizes (when is displayed), press
[ 0/T]
The display will show zero.
4. Place the sample in the tare container
(1) Open the glass draft shield.
(2) Place the sample on the pan, and then close the glass draft shield.
5.
When the display stabilizes (when is displayed), read the
result on the display
*
*