EasyManua.ls Logo

Shimadzu AP W-AD Series - Measuring Averages (Averaging mode)

Shimadzu AP W-AD Series
169 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
95
1
3
5
2
4
6
7
8
12
14
13
9
10
11
77 ApplicationsApplications
Measuring Averages (Averaging mode)
Makes it possible to restrain fluctuations of the weighing value when weighing moving samples such
as small animals and when weighing in bad conditions with strong vibrations, etc. The hold function
can also be used for measuring averages. This enables the average value of the measurements to
be calculated for a predetermined period of time when it has been set at ON. Averaging mode is
performed in the real-time when the hold function is set at OFF.
Preparing for Averaging modes
Perform the preparations required for carrying out Averaging mode. Set the hold function to ON or
OFF.
1. Set the Averaging mode
(1) Press [ MENU] while in the weight
measurement mode, and then press [ LEFT].
(2) Select [
Select Mode], and then press [ OK].
(3) Select [
Averaging mode
], and then press
[ OK].
(4) The measurement screen will be displayed.
2. Start the Averaging mode settings
Press [ MENU].
3. Set the hold function ON or OFF
(1) Select [Hold], and then press [ OK].
(2) Select [on] or [off], and then press [ OK].
· Proceed to Step 4 if [on] was selected.
· Commence Averaging modes if [off] was
selected.
4. Set the measurement time
(1) Select [Setting] and then press [ OK].
(2) Select [Measuring time], and then press [ OK].
(3) Enter the measurement time between 1 and 999
seconds, and then press [ OK].
^ [Entering Numerals and Characters] (P.43)
What is measurement time?
This is the period of time required for measurements
after the sample has been placed on the pan and the
[ OK] button pressed. The weight will be
measured and the average value calculated during
this period.
*