Adjusting the Stability Mark
The stability mark is displayed when the weighing value stabilizes.
The following settings are available as the conditions for illuminating (Stability Mark),
· Stability detection range
· Stability mark ON timing
Although it is not necessary to change this setting under normal circumstances, it does need to be
changed if the stability judgment conditions are to be weakened (stability mark) so that detection
is easier when the balance is installed in an unstable environment, for example.
It is recommended that the [Stability detection range] and [Stability mark ON timing] parameters are
adjusted in accordance with the way in which the balance is being used while referring to the following
chart in order to set the appropriate stability detection level.
Prioritizing operability:
Prioritizes operability even in easily-changeable
measurement environments to display the
Stability Mark swiftly.
Acquiring reliable data:
It takes time for
the Stability Mark to be
displayed, but the weight is stable after it has
been displayed.
Fast
Standard
Stability mark ON
timing
Stability detection range (Count)
Stability detection range Settings
The Stability detection range determines whether the weighing value is stable or not by judging if the
value is within the preset count for the minimal digit for a predetermined period of time.
(The default for the Stability detection range is set a 1 count (1d).)
1. Call out the measurement setting menu
(1) Press [ MENU] while in the weight measurement mode, and then press [ LEFT].
(2) Select [ Measurement Setting], and then press [ OK].
2. Set the Stability detection range
(1) Select [Stability detection range], and then press
[ OK].
(2) Select the value for the Stability detection range,
and then press [ OK].
Select either 0.5, 1, 5, 10, 50, 100 or 1000 for the
Stability detection range in alignment with the use
or objective of the measurement.
3. Return to the weight measurement mode
Press [ POWER]
When data output is slow
The installation environment and the sample are the reasons for unstable displays. When the
data output related to stability inspections is slow, increase the Stability detection range.
*