77 ApplicationsApplications
Changing the Weight of the Weight
Press [ MENU], enter the weight of the weight and
then press [ OK] to change the weight value.
(4) Remove the weight in accordance with the screen
instructions, and then press [ OK].
(5) Press [ 0/T].
(6) Repeat procedures from (3) to (5) in accordance
with the screen instructions.
(7) Read the Pass/fail evaluation once minimum
weighing value measurement has been completed.
Minimum weighing value setting
If the minimum weighing value has been actually
measured and has been judged as [Pass]:
Press [ OK]: Updates the minimum weighing value setting.
Press [ POWER]: The minimum weighing value is
not updated. If the value has been judged as [Fail], check
the suitability of the balance’s installation environment,
and then increase the weight and measure it again.
The displayed value maybe changed by your balance model.
5.
Return to the weight measurement mode
Press [ POWER].
Minimum weighing value output
Note that it is not possible to print or output a value that is lower than the minimum weighing
value when in the minimum weighing value mode.
Setting a random value for the minimum weighing value
Observe the following procedure if the minimum weighing value
has been measured and a value larger that this is to be reset in the
balance. However, this function cannot be used unless the minimum
weighing value has been actually measured and calculated.
1. Setting the minimum weighing value
(1) Select [Arbitrary setting], and then press [ OK].
(2) Enter a random minimum weighing value, and then
press [ OK].
^ [Entering Numerals and Characters] (P.43)
2.
Set an arbitrary minimum weighing value.
(1) Select [Arbitrary Setting], and then press [ OK].
(2) Enter the arbitrary minimum weighing value, and then press [ OK].
^ [Entering Numerals and Characters] (P.43)
3. Return to the weight display.
Press [ POWER].
Setting minimum weighing values for USP (United States Pharmacopeia)
The minimum weighing value is not set when the balance in shipped from the factory. It is not
possible to set a value that is 820-times lower than the minimum weighing value listed in the
USP (United States Pharmacopeia). It is also not possible to set a minimum weighing value that
is lower than the value acquired through actual measurement.
Service setting
The [Service Setting] shown in the above-mentioned [MW Judgment] menu is used by representatives from
the SHIMADZU service company who visits the user’s premises to set the minimum weighing value. This menu
cannot be operated by users. Contact the SHIMADZU service company for further details on the value set.
*
When the minimum weighing value has been
judged as [Pass]
When the minimum weighing value has been
judged as [Fail]
*
*
*
*
*

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu AP W-AD Series and is the answer not in the manual?

Shimadzu AP W-AD Series Specifications

General IconGeneral
BrandShimadzu
ModelAP W-AD Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals