55 CalibrationCalibration
Time-Specied Adjustment (Timer CAL) Settings (W-AD/ W/ X Series only)
It is possible to set the timer and perform PSC (Perfect Self Calibration) and test at a predetermined
time with the built-in weight in W series models. Three different times can be registered. This function
is useful when calibration records are required for regular sensitivity adjustment, and when it is
necessary to adjust sensitivity during break periods, etc., so as not to interfere with measuring work.
The weight mark will start blinking approximately two minutes prior to sensitivity adjustment to
notify the user.
Time-specified adjustment outside of weight display modes
Time-specified adjustment are deactivated outside of the stand-by mode, the menu display
mode, the input mode and other Weight display modes.
1. Call out the calibration menu
Press [ CAL] for three or more consecutive
seconds.
2. Select the timer to set
(1) Select [Timer CAL], and then press [ OK].
(2) Either any of the timers from 1 to 3, and then press
[ OK].
3. Set the timer ON/OFF
When configuring [Timer CAL] for the first time,
proceed to step 4.
To enable or disable an already configured timer
select [on] or [off] then press [ OK].
4. Set the time
(1) Select [Time setting], and then press [ OK].
(2) Enter the time for the timer, and then press
[ OK].
^ [Entering Numerals and Characters] (P.43)
5.
Return to the weight measurement
mode
Press [ POWER]
In case of using the Multi-stand or AP Holder
Remove the Multi-stand or AP Holder from pan, or set off the timer-CAL function.
When configuring the time setting, set the time to a time at least 2 minutes after the time while
configuring the settings. If set to less than 2 minutes, the timer may not activate even when the
specified time is reached.
*
*

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu AP W-AD Series and is the answer not in the manual?

Shimadzu AP W-AD Series Specifications

General IconGeneral
BrandShimadzu
ModelAP W-AD Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals