8015416/ZPN2/2017-09-08 • © SICK AG • Subject to change without notice 123
Operating instructions Distance measuring device DL100 – CANopen®
Troubleshooting
Display Meaning / possible causes Troubleshooting
eLevel Current damping value is below the warning
threshold. The warning threshold depends
on the distance between measuring device
andreector.For recommended damping
values, see page 32, Table 4.
Clean external lens surfaces like the
reectorandthelens.
Decrease the distance between the
measuringdeviceandthereector.
Use a distance measuring device with a
higher range.
See page 126, chapter 14.2.
eTemp The internal device temperature is outside of
the permissible range.
For the permissible ambient temperature,
see page 128, chapter 14.9.
Check ambience temperature, improve
ventilation if applicable.
Shield against radiation heat, e.g. shade
the measuring device in case of direct
solar irradiation.
Use device with heating at low ambient
temperatures.
Use cooling housings for high ambient
temperatures.
Table 76: Error messages
12.4 Return
Forecientprocessingandquickdeterminationofcauses,includethe
following in your return:
Information on a contact
A description of the application
A description of the error that occurred
12.5 Disposal
Observe the following items for disposal:
The distance measuring device must not be disposed of in the house-
hold waste.
Dispose of the distance measuring device according to the respective
country-specicprovisions.