8015416/ZPN2/2017-09-08 • © SICK AG • Subject to change without notice 51
Operating instructions Distance measuring device DL100 – CANopen®
Operation at the measuring device
8.5.10 Submenu „MF1 – Srvice“
This submenu is used to parameterize the multifunction output MF1 as
serviceoutput.Youmayactivate(on)ordeactivate(o)severaloptions.
You can get to the menu "Srvice" via the menu path:
Main menu →
→ Menu → → CANopn → → more → →
MFx On → → MF1 → → ActSta → → Functn → → Srvice
Requirements for the display
• Menu "more": Option "Yes"
• Menu "MFx On": Option "Enable"
• Parameter "Functn": Option "Srvice"
Parameter Description
WrnLsr Activating and deactivating warning messages when the mea-
suring device must be replaced soon because the laser ages.
Options
• On
• O
Factory setting
• On
WrnLvl Activate or deactivate warning messages when the damping
value is undercut, e.g. at contamination.
Options
• On
• O
Factory setting
• On
WrnTemp Activate or deactivate warning message when the inner tem-
perature of the measuring device is outside of the permissible
thresholds.
Options
• On
• O
Factory setting
• On
WrnPlb Activate or deactivate warning when the measurement value
is not plausible. Possible reasons may be incorrect measure-
ments, interruption of the light beam, optical interferences or
electrical interferences.
Options
• On
• O
Factory setting
• On