Operating instructions Distance measuring device DL100 – CANopen®
Technical data
14.4 Performance
Measurement ranges DL100–11XXXXXX: 0.15 m … 60 m
DL100–12XXXXXX: 0.15 m … 130 m
DL100–13XXXXXX: 0.15 m … 220 m
Measuring accuracy Seetypespecicdata page 126, chapter 14.2
Repeatability Seetypespecicdata page 126, chapter 14.2
Initialization time Typical 1.5 s
Afterreectorloss:<40ms
Reaction time/Dead time 2 ms
Resolution Adjustable: 0.1 mm / 0.125 mm / 1.0 mm / 10 mm / 100 mm / free
Output rate synchronous to PLC request
Table 79: Performance data
14.5 Supply
Supply voltage VS 18 V DC … 30 V DC
Current consumption Without heating: < 250 mA at 24 V DC
With heating : < 1.000 mA at 24 V DC
Residual ripple < 5 Vss within the permissible supply voltage UV
Table 80: Supply
14.6 Inputs
Inputs Multifunction input MF1, adjustable
Hi>12V
Lo < 3 V
See page 48, Table 15, parameter “ActSta”.
Protective circuit No, not reverse polarity protected
Table 81: Inputs