Operating instructions Distance measuring device DL100 – CANopen®
Operation at the measuring device
56 © SICK AG • Subject to change without notice • 8015416/ZPN2/2017-09-08
Menu “SpecFu”
(continued)
Parameter Description
AvgSpd Selectlterdepthforthespeedvalues.
Options
• Medium
• Slow
• Fast
Factory setting
• Medium
ErrRej Select time for error suppression. During this time, the old
measurement value is output. When there still is no valid
measurement value after the time selected for the parameter
"ErrRej", the value "0" is output.
Options
• 200ms: Error/warning is indicted when the error is present
for longer than 200 ms.
• 50ms: Error/warning is indicted when the error is present
for longer than 50 ms.
• OError/warningisindicatedatonce,withoutdelay.
Factory setting
• 200ms
Heat Requirements for the display
• Only for versions with heating DL100-xxHxxxxx
This menu is used to set the temperature at which the heating
istoactivate.Thehysteresisissetrmlyto2K.
Adjustment range
• -10…+40°C
Factory setting
• -10 °C
FMode Requirements for the display
• Only for versions with frequency switching DL100-xxxBxxxx
Select frequency range Frequency switching may be required
at parallel placement of several distance measuring devices.
→ also see page 29, chapter 6.4.
Options
• Mode 1
• Mode 2
• Mode 3
• Mode 4
Factory setting
• Mode 1
Reset Perform reset → see page 57, chapter 8.6.
Table 22: Menu “SpecFu”