9. Remove the two socket head cap screws (P/N 0004-645-000) (BT) that hold the IR
board bracket (0702-001-326) (AY) to the rover chassis using a 3/16” allen wrench.
The IR board bracket assembly can remain connected while removing the canister.
. Remove the two socket head cap screws (P/N 0004-645-000) (BT) that hold the IR
board bracket (0702-001-326) (AY) to the rover chassis using a 3/16” allen wrench.
The IR board bracket assembly can remain connected while removing the canister.
Caution: When removing the inlet hose in step 10, be aware that
water and detergent may be present in the hose!
10. Remove the coupling clip (P/N 702-1-204) (C) from the coupling block (P/N 702-1-
600) that holds the diverter inlet water hose assembly (P/N 702-1-341) (B) and
remove the hose from the block.
10. Remove the coupling clip (P/N 702-1-204) (C) from the coupling block (P/N 702-1-
600) that holds the diverter inlet water hose assembly (P/N 702-1-341) (B) and
remove the hose from the block.