2. Disconnect the vacuum pump wire harness (P/N 0702-001-819) from the right side of
the vacuum pump.
3. In order to allow enough clearance for the vacuum pump to be removed, you may have
to remove the silencer manifold assembly (P/N 0702-001-160) (J). If this is required,
loosen the flare swivel adapter that connects the silencer to the vacuum pump using a
7/8” open-ended wrench.