☺ Starting with MultiFlow software version 3.30[3.30] two cards of type CS-
CC-64 are required; therefore after approx. 60 seconds the following
message is displayed:
„Please insert next Chip Card!“
Confirm the change of the chip card with <F1> („OK“).
☺ When the data transfer is completed the following message is displayed:
"Data transfer complete!"
Acknowledge this message with <F1> ("OK").
☺ The parameters are now transferred.
If an error occurs, an appropriate error message is given and the transfer
needs to be repeated.
9.12.2
Loading the Parameters from the Chip Card
Data is read from a parameter chip similarly to the saving procedure in
menu 4.5.1.
Only deviating memory ranges are transferred!
However, here particular attention should be given to the fact that in the
sealed state the MultiFlow only accepts parameters which are not
subject to W & M regulations.
If the parameters subject to calibration regulations deviate from the
present settings, the following instruction is given:
"Break the electr. seal before processing this parameter!"
Acknowledge this message with <F1> ("OK").
Then the following message appears:
"Data transfer unsuccessful!"
The error message only applies to the transfer of parameters subject to
calibration regulations. All others will have been transferred.
☺ Starting with MultiFlow software version 3.30[3.30] two cards of type CS-
CC-64 are required; therefore after about. 60 seconds the following
message is displayed:
„Please insert next Chip Card!“
Confirm the change of the chip card with <F1> („OK“).