5.3 Reference Guide using RS-232C commands
138
READ:RFTC:ES:CH?
Description
This command reads channel value for EDR Sensitivity(RCV/CA/07/C) measurement.
CONF:RFTC:ES:CH <Value>
Description
This command sets channel for EDR Sensitivity(RCV/CA/07/C) measurement.
Value
channel : 0-39-78(default)
(ex) In case of setting channel 0 : CONF:RFTC:ES:CH 0
In case of setting channel 0, 39, 78 : CONF:RFTC:ES:CH 0-39-78
READ:RFTC:ES:PACK_TYPE?
Description
This command reads packet type for EDR Sensitivity (RCV/CA/07/C) measurement
CONF:RFTC:ES:PACK_TYPE <Value>
Description
This command sets packet type for EDR Sensitivity(RCV/CA/07/C) measurement
Value
Packet type : 2-LONGEST(default),3-LONGEST,2-DH1,3-DH1,2-DH3,3-DH3,2-DH5,3-DH5,2-EV3,3-EV3,2-
EV5,3-2V5
READ:RFTC:ES:HOP?
Description
This command reads hopping mode for EDR Sensitivity(RCV/CA/07/C) measurement.
CONF:RFTC:ES:HOP <Value>
Description
This command sets hopping mode for EDR Sensitivity(RCV/CA/07/C) measurement.
Value
Hopping Mode : OFF(default), ON
READ:RFTC:ES:DIRTY_TX?
Description
This command reads Dirty Transmitter value for EDR Sensitivity(RCV/CA/07/C) measurement.
CONF:RFTC:ES:DIRTY_TX <Value>
Description
This command sets Dirty Transmitter for EDR Sensitivity(RCV/CA/07/C) measurement.
Value
Dirty Transmitter Mode : OFF, ON(default)
READ:RFTC:ES:RX_LEV?
Description
This command reads DUT RX Level value for EDR Sensitivity(RCV/CA/07/C) measurement.
CONF:RFTC:ES:RX_LEV <Value>
Description
This command sets DUT RX Level for EDR Sensitivity (RCV/CA/07/C) measurement.
Value
DUT RX Level : -327~20, -70(default)