5.3 Reference Guide using RS-232C commands
139
READ:RFTC:ES:BYTES?
Description
This command reads number of received byte for EDR Sensitivity(RCV/CA/07/C) measurement.
CONF:RFTC:ES:BYTES <Value>
Description
This command sets number of received byte for EDR Sensitivity(RCV/CA/07/C) measurement.
Value
DUT RX Level : 0 ~ 2147000000, 200000(default)
READ:RFTC:ES:BER?
Description
This command reads BER value from EDR Sensitivity(RCV/CA/07/C) measurement. Before reading this
value use “READ:RFTC:ES:BER?” Command, RF test case should be executed to update this value by
sending “EXEC:RFTC:ES:RUN”.
5.3.24 EDR BER Floor Performance(EBP)
EXECRFTC:EBP:RUN
Description
This command executes EDR BER Floor Performance(RCV/CA/08/C) measurement.
EXEC:RFTC:EBP:STOP
Description
This command stops exection of running EDR BER Floor Performance(RCV/CA/08/C) measurement. It is
used to cancel the running RF test case which is started by “EXE:RFTC:EBP:RUN” command.
READ:RFTC:EBP:CH?
Description
This command reads channel value for EDR BER Floor Performance(RCV/CA/08/C) measurement.
CONF:RFTC:EBP:CH <Value>
Description
This command sets channel for EDR BER Floor Performance(RCV/CA/08/C) measurement.
Value
channel : 0-39-78(default)
(ex) In case of setting channel 0 : CONF:RFTC:EBP:CH 0
In case of setting channel 0, 39, 78 : CONF:RFTC:EBP:CH 0-39-78
READ:RFTC:EBP:PACK_TYPE?
Description
This command reads packet type for EDR BER Floor Performance(RCV/CA/08/C) measurement.
CONF:RFTC:EBP:PACK_TYPE <Value>
Description
This command sets packet type for EDR BER Floor Performance(RCV/CA/08/C) measurement.
Value
Packet type : 2-LONGEST(default),3-LONGEST,2-DH1,3-DH1,2-DH3,3-DH3,2-DH5,3-DH5,2-EV3,3-EV3,2-
EV5,3-2V5
READ:RFTC:EBP:HOP?