2.6 Cleaning, Storage and Shipment
94
4.3.2 Maximum sensitivity
Specification: Maximum value <= -100 dBm @ resolution BW = 40 kHz
1. Test Setup: Connect equipment as shown in the figure below
[Figure 11] Maximum sensitivity test
2. Set controls of TC-3000C as follows:
Make sure “RX Attenuator” is 0dB. ( MENU Select “Configuration” M3 F3 )
Set TC-3000C to ”Spectrum” screen mode. ( MENU Select Measurement”
M1 )
- Frequency: 2450 MHz
- Ref. Level: -50 dBm
- Span: 10 MHz
- Num of sweeps: 50
3. Read the peak level on the screen.
TC-3000C
50 Terminator