2.6 Cleaning, Storage and Shipment
88
4.2.2 I/Q modulation calibration
Equipment: Spectrum Analyzer, HP8561E
1. Test Setup: Connect equipment as shown in the figure below
[Figure 6] I/Q modulation calibration test
2. Set controls of TC-3000C as follows:
Set TC-3000C to “Signal Generator” mode. ( MENU Select “SigGen”)
- Power: 0dBm
- Bit Patten: 0
- FM Deviation:150 kHz
3. Set controls of Spectrum Analyzer as follows:
Span: 1 MHz
RBW: 10 kHz
VBW 210 kHz
4. Set TC-3000C output frequency according to the table below. At each setting,
press “Start” ( F1 ) to begin the test.
5. Record the level in the table for each setting. The limits for this frequency are
given in the table.