Appendix B. RF Test Cases of the TC-3000C
216
selected then the displayed value of the parameter is ignored and the value according to RF
Test Specification [1] is assigned automatically.
Requirement to Initial Condition of a DUT
If the DUT can be controlled through HCI, the DUT must be connected to an appropriate HCI
port of the tester and correct value of the HCI Port parameter must be specified in DUT
configuration (MenuConfigurationDUTHCI Port).
If the DUT does not have HCI then the DUT must be in the Inquiry Scan/Page Scan state
(maximum activity of the scan mode is desirable) and Test Mode should be enabled (for
Transmitter Test and Loop back Test Modes).
5.2.3 Power Control (TRM/CA/03/C)
Goal
Verification of the TX power control
Parameters
1. Test mode: Transmitter test, Loop back, Null packets (simple connection).
2. BT channels are the TX frequency channels of the DUT, which are selected to test. The
frequency for each channel is calculated as (2402+ch) MHz, where ch is the frequency
channel. The frequency channel is the value from 0 to 78. Any quantity of the frequency
channels can be selected in the TC-3000C. The list of selected channels must be
separated by “-“. For instance, the record “0-20-40-60” means that the test case must be
executed for four frequency channels: 0, 20, 40 and 60, that is for the frequencies
2402MHz, 2422MHz, 2442MHz and 2462MHz.
3. Packet type parameter specifies which packet type will be used for the test case. If the
parameter is equal to ‘longest’ then the tester will read LMP features of the DUT and the
longest supported packet will be chosen for the test case.
Note. This parameter is ignored in Null packets test mode.
4. Number of packets parameter specifies how many packets will be used to find the
average RF-output power for each power step.
5. Measurement method specifies the algorithm, which will be used for the test case. If the
parameter is equal to 0 then the algorithm, described in [1] is used. If the parameter is
equal to 1 then fast algorithm is used.
If the flag “Use Default” is selected then all of displayed values of parameters are ignored and
the listed above values will be assigned automatically:
Test/Normal Mode: Transmitter
BT channels: 0-39-78
Packet type: DH1
Number of packets: 1
Measurement method: 0 (BT Spec)
Additional test conditions
In Transmitter test and Loop back test mode whitening is disabled, the payload’s bit pattern is
PRBS9. If the Measurement method = 0, then in Transmitter test and Loop back test mode the