5.3 Reference Guide using RS-232C commands
125
READ:RFTC:SM:DIRTY_TX?
Description
This command reads dirty transmitter mode.
CONF:RFTC:SM:DIRTY_TX <Value>
Description
This command writes dirty transmitter mode.
Value
dirty transmitter mode : ON(default), OFF
READ:RFTC:SM:BER?
Description
This command reads bit error rate.
5.3.16 Maximum Input Level(ML)
EXEC:RFTC:ML:RUN
Description
This command executes measurement of maximum input level test case(RCV/CA/06/C).
EXEC:RFTC:ML:STOP
Description
This command stops execution of maximum input level.
READ:RFTC:ML:CH?
Description
This command reads RX frequency of a tester(TX for a DUT).
CONF:RFTC:ML:CH <Value>
Description
This command writes RX frequency of a tester(TX for a DUT).
Value
channel : RX frequency of a tester (TX for a DUT), 0 ~ 78, 0-39-78(default)
READ:RFTC:ML:HOP?
Description
This command reads hopping mode.
CONF:RFTC:ML:HOP <Value>
Description
This command writes hopping mode.
Value
hopping mode : ON, OFF(default)
READ:RFTC:ML:RX_LEV?
Description
This command reads RX power at the receiver input of the DUT in dBm.
CONF:RFTC:ML:RX_LEV <Value>
Description
This command writes RX power at the receiver input of the DUT in dBm.