5.3 Reference Guide using RS-232C commands
140
Description
This command reads hopping mode for EDR BER Floor Performance(RCV/CA/08/C) measurement.
CONF:RFTC:EBP:HOP <Value>
Description
This command sets hopping mode for EDR BER Floor Performance(RCV/CA/08/C) measurement.
Value
Hopping Mode : OFF(default), ON
READ:RFTC:EBP:DIRTY_TX?
Description
This command reads Dirty Transmitter value for EDR BER Floor Performance(RCV/CA/08/C) measurement.
CONF:RFTC:EBP:DIRTY_TX <Value>
Description
This command sets Dirty Transmitter for EDR BER Floor Performance(RCV/CA/08/C) measurement.
Value
Dirty Transmitter Mode : OFF(default), ON
READ:RFTC:EBP:RX_LEV?
Description
This command reads DUT RX Level value for EDR BER Floor Performance(RCV/CA/08/C) measurement.
CONF:RFTC:EBP:RX_LEV <Value>
Description
This command sets DUT RX Level for EDR BER Floor Performance(RCV/CA/08/C) measurement.
Value
DUT RX Level : -327~20, -60(default)
READ:RFTC:EBP:BYTES?
Description
This command reads number of received byte for EDR BER Floor Performance(RCV/CA/08/C)
measurement.
CONF:RFTC:EBP:BYTES <Value>
Description
This command sets number of received byte for EDR BER Floor Performance(RCV/CA/08/C) measurement.
Value
Number of samples : 0 ~ 2147000000, 1000000(default)
READ:RFTC:EBP:BER?
Description
This command reads BER value from EDR BER Floor Performance(RCV/CA/08/C) measurement.
Before reading this value use “READ:RFTC:EBP:BER?” Command, RF test case should be executed to
update this value by sending “EXEC:RFTC:EBP:RUN ”.
5.3.25 EDR Maximum Input Level(EML)
EXEC:RFTC:EML:RUN
Description
This command executes EDR Maximum Input Level(RCV/CA/10/C) measurement.