3.3 Operation Procedure (Link Analyzer)
40
2. Setting up # Inq. Response: You can set the Inquiry to end after the “# Inq.
Responses” (1 to 16). If there are several DUT that have to respond to Inquiry, set the
number of Inquiry response to more than 1. But if there is only one DUT, set the
number of Inquiry response to 1 for testing speed. TC-3000C will stop discovering
when receives this number of inquiry responses or the Inquiry Timeout is expired.
Press F8 one times for “More 2/4” menu Press F4
3. Search for DUT: Press F8 three times for More 1/4” menu press F2 key to
search for all Bluetooth devices that are in range and in Inquiry Scan Mode through
the Inquiry procedure. If the DUT is connected to the TC-3000C through the HCI port,
the TC-3000C read the Bluetooth device (BD) address and controls the DUT directly
through USB(HCI). If you want to stop Inquiry process, press F2 key again.
4. Select a DUT: Select a DUT that you want to test in the found devices by using
Rotary Knob or
keys.
5. Connect/Disconnect: Create connection to highlighted DUT on the screen through
the Page procedure by pressing F3 . When the highlighted DUT is connected, this
menu changes to Disconnect. If you want make a SCO connection, press F4 .
6. Set DUT to Test Mode: Enable DUT to the test mode by pressing F5 . If DUT is not
connected, the page procedure will be performed. If there is no DUT found, the
discovering (inquiry) procedure will start first.
Note
If you want to edit test configuration in detail, you can access to configuration
menus by pressing F8 in sequence. These menus operate globally; changing the
settings in this screen automatically changes that setting in all screens where it is
available. Refer to “3.9 Configuration”
[Inquiry Parameters ]
Parameters
Range
Default
Descriptions
Inq. TO (1.28s)
1 ~ 48
48
Refer to Tester in Configuration
Page TO (slts)
1 ~ 65535
16000
Refer to Tester in Configuration
# responses
0 ~ 16
1
Refer to Tester in Configuration
HCI TO (ms)
1 ~ 65535
2000
Refer to Tester in Configuration
[ Scan Parameters ]
Parameters
Range
Default
Descriptions
Access Type
Discoverable
Connectable
Full Access
Full Access
Accessibility to Tester as a slave
T inqscan (slts)
18 ~ 4096
2048
Refer to Tester in Configuration
T winqscan (slts)
18 ~ 4096
18
Refer to Tester in Configuration
T pagescan (slts)
18 ~ 4096
2048
Refer to Tester in Configuration
T wpagescan (slts)
18 ~ 4096
18
Refer to Tester in Configuration
Scan TO (ms)
10000 ~
600000
Inquiry/Page Scan Timeout of Tester
[Table 4] Test Parameters Refer to DUT in Configuration.