2.6 Cleaning, Storage and Shipment
91
4.2.5 Output level accuracy test
Equipment: Spectrum Analyzer, HP8561E
Signal Generator, HP 8648C
Mixer
Power Sensor
Specification: Power error <= ±1 dB
1. Test Setup: Connect equipment as shown in the figure below
[Figure 9] Output level accuracy test
2. Set controls of TC-3000C as follows:
Set TC-3000C to “Signal Generator” mode. ( MENU Select “SigGen”)
- Frequency: 2400 MHz
- Power: 0dBm
- Bit Patten: 0
- FM Deviation: 0
3. Connect the power sensor to TC-3000C
4. Set TC-3000C output frequency and level according to the table below. At each
setting, press “Start” ( F1 ) to begin the test.
5. Record the power in the table for each setting.