EVM Reports and Printing
EVM and Security Settings
EVM Series User Manual
EVM and Security Settings
The security or lock feature is used to prevent a user from interrupting a run and/or prevents someone
from changing a setup or saved setups on the instrument.
NOTE: A default code, “9157”, will disable secure run or secure setup if the numeric Passcode is
forgotten.
1. Click on the Security tab (see ❶) from the EVM’s Instrument Communications panel.
2. To set the Secure Setup Mode or Secure Run Mode, follow below:
Click in the Enabled box to turn on secure run setup mode and/or secure run mode.
3. Type in a passcode (4 numeric characters).
4. To set the auto-run, click on or save and send to the EVM (for more details see
“Saving and sending configurations”).
Figure 6-13: EVM Auto-Run with Day of Week setting
EVM and Triggering Settings
There are two types of triggering outputs which may be programmed to with one of the seven sensors.
NOTE: Refer to the EVM triggering table identifies the components of this setup screen.
Digital output: You can connect an external device to the monitor. This can be used to signal a specific
level to turn on a notification device, an LED, or a fan motor at a specific threshold (or set point). Once
enabled, the following three are selectable parameters: a sensor, logic setting, and threshold setting.
Analog output: Enables the analog output on the device when checked. The analog output will vary
between 0–5 volts based upon the minimum and maximum levels for the particular sensor.
1. Click on the Triggering tab (see ❶) from the EVM’s Instrument Communications panel.