163 Appendix E
System Errors Displayed on the EVM Series
EVM Series User Manual
Appendix E
System Errors Displayed on the EVM Series
There are two types of pop-up notifications that may appear on your instrument.
Monitor system errors – The EVM is equipped with numeric system errors which range from
101 to 108. If a numeric system error appears, please call TSI for Technical Support and Service.
Noting the System Error code is very useful for the call.
Monitor errors – Pop-up screens may appear as you are using the instrument. These are to notify you
of various errors/cautions statements. All of the errors are listed below with an explanation. Press Enter
and adhere to the monitor warning screens. If you have any questions, please call TSI for
Technical Support.
Monitor Errors
Explanation
Battery charger disabled at
current temperature
The battery is either too hot or too cold to charge.
Battery charging failure
If this appears frequently, this indicates the instrument may
need to be serviced. Contact TSI’s Technical Support for help.
Battery low
The battery is almost depleted and powering off of external
power.
NOTE: Turn the monitor off or plug in the power cord.
Right error: Duplicate profile
name
The dust profile names that the user can set must be unique or
this error will appear. (See Chapter 3 for details.)
Analog-out channel sensor invalid
If analog-out is enabled and a sensor is selected (such as CO)
and you remove that sensor, this monitor warning will appear.
(Re-insert the sensor if desired.)
NOTE: The Digital-out field will automatically disable and select
the next available detectable sensor.
Invalid impactor setting to start
logging
When the impactor reads at an incorrect setting, it will not allow
you to log without turning the turret to a correct setting.
Laser not functioning properly
Either the laser is unplugged or needs to be serviced. Make
sure the laser is connected (This is the multi-colored cable
which attaches from the laser to the connectors located by the
gravimetric cassettes.)
Logging stopped due to impactor
change
The instrument does not allow you to change the impactor
setting during run mode. To correct this, stop logging. Change
the impactor and then continue with Running/logging
measurements.
APPENDIX
E