3-16 Tuning the Instrument
The Instrument Setup Wizard Lock Mass check is complete window
appears.
Instrument Setup Wizard Lock Mass check is complete:
12. Ensure that the reference reservoir valve is closed.
If you have injected more sample or backed sample out using the
reference reservoir valve, the signal can overshoot and appear too
intense or too weak.
13. Click Back to repeat the Lock Mass intensity check.
14. Click Next.
The Instrument Setup Wizard DRE Magnification Factor Setup window
appears.