5-35
IM 701240-01E
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
App
Index
Horizontal and Vertical Axes
Frequency
The procedure after selecting Frequency in step 8 on page 5-34 is explained below.
9. As necessary, use jog shuttle+SELECT to enter the smoothing filter, pulse
average, deceleration prediction, and stop prediction.
Smoothing filter
Moving average order of smoothing
Stop prediction
Deceleration prediction
Pulse average
Pulse average count
Setting the Measurement Range
10. Turn the RANGE knob to set the measurement range.
You can also set the measurement range by pressing the Range soft key and turning the
jog shuttle.
RPMs/RPSs
The procedure after selecting Revolution (rpm) or Revolution (rps) in step 8 on page 5-
34 is explained below.
9. Use jog shuttle+SELECT to set the number of pulse per rotation (Pulse/
Rotate).
10. As necessary, use jog shuttle+SELECT to enter the smoothing filter, pulse
average, deceleration prediction, and stop prediction.
Smoothing filter
Moving average order of smoothing
Stop prediction
Deceleration prediction
Pulse average
Pulse average count
Number of pulses per rotation
Setting the Measurement Range
11. Turn the RANGE knob to set the measurement range.
You can also set the measurement range by pressing the Range soft key and turning the
jog shuttle.
5.17 Setting the Frequency (Number of Rotations, Period, Duty Cycle, Power Supply Frequency, Pulse Width,
Pulse Integration, and Velocity) Measurement

Table of Contents