8-14
IM 701240-01E
Setting the Display Range
10. Press the Start Point/End Point soft key to set the jog shuttle control to Start
Point.
11. Turn the jog shuttle to set the start point of the X-Y waveform display.
12. Likewise, press the Start Point/End Point soft key to set the jog shuttle control
to End Point and set the end point of the X-Y waveform display.
Explanation
X-Axis Mode
Single: The X-axis is set to a common trace, and the Y-axis displays all the waveforms
whose display is turned ON.
Quad: The X-axis trace and Y-axis trace are set individually for XY1 to XY4.
Assigning X (Horizontal) and Y (Vertical) Axes (When X Axis Is Set to
Quad)
The channels that are assigned to the X and Y axes depending on the X-axis mode are
as follows:
X-Axis Mode X-Y Waveform X-Axis Y-Axis
Single CH1 to CH16 and All channels that are turned
Math1 to Math8 ON
Quad XY1 CH1 to CH8, Math1 to CH1 to CH8, Math1 to
Math4 Math4
XY2 CH1 to CH8, Math1 to CH9 to CH16, Math5 to
Math4 Math8
XY3 CH9 to CH16, Math5 to CH1 to CH8, Math1 to
Math8 Math4
XY4 CH9 to CH16, Math5 to CH9 to CH16, Math5 to
Math8 Math8
Number of X-Y Waveforms That Can Be Displayed
The number of X-Y waveforms that can be displayed is 16 for Single and 4 for Quad.
The display of each X-Y waveform is turned ON/OFF by turning the Y-axis trace display
ON/OFF.
Display Mode
You can select from the following three display modes.
T-Y&X-Y: The top window displays T-Y (normal) waveforms. The bottom window
displays X-Y waveforms.
X-Y: Displays only X-Y waveforms.
T-Y: Displays only T-Y waveforms.
8.6 Displaying X-Y Waveforms

Table of Contents