AutoTest Sytem Operation
Subject to Export Control, see Cover Page for details.
5-28
Set Commands (cont)
set_analyzer_marker1topeak
Moves marker to peak point on trace.
Must insert a 1 s wait command when this command follows the
set_analyzer_marker1enable command or set_analyzer_marker1position command.
set_analyzer_marker1tomin
Moves marker to lowest point on trace.
set_analyzer_marker1setscf
Sets Center Frequency to frequency at defined marker position.
set_analyzer_marker1setsref
Sets Reference Level to level at defined marker position.
set_analyzer_marker1setsright
Moves marker right to next peak point on trace.
set_analyzer_marker1setsleft
Moves marker left to next peak point on trace.
set_analyzer_marker1_ulimit [ limit dBm ]
Sets analyzer marker upper limit.
Format: Decimal limit value in dBm
set_analyzer_marker1_ulen [ state 0/1 ]
Enables/Disables the upper limit check on Marker 1.
Valid states are:
0 = Disabled
1 = Enabled
set_analyzer_marker1_llimit [ limit dBm ]
Sets analyzer marker lower limit.
Format: Decimal limit value in dBm
set_analyzer_marker1_llen [ state 0/1 ]
Enables/Disables the lower limit check on Marker 1.
Valid states are:
0 = Disabled
1 = Enabled
set_analyzer_marker2enable [ state 0/1 ]
Enables/Disables Marker at set position.
Valid states are:
0 = Disabled
1 = Enabled
set_analyzer_marker2position [ value in MHz ]
Sets marker position.
Format: Decimal value in MHz