EasyManuals Logo

Agilent Technologies 4288A User Manual

Agilent Technologies 4288A
289 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #217 background imageLoading...
Page #217 background image
Appendix B 215
Information for Replacing 4278A with 4288A
Functional Comparison and GPIB Command Correspondence
B. Information for Replacing
4278A with 4288A
GPIB command correspondence
Table B-4 Correspondence of measurement range setup commands
Setup
GPIB command
Measurement
frequency
Measurement range
4278A 4288A 4278A 4288A
1 kHz
Auto Auto RA0 :RANG:AUTO ON
100 pF 100 pF RA1 :RANG 100PF
1 nF
220 pF
RA2
:RANG 220PF
470 pF :RANG 470PF
1 nF :RANG 1NF
10 nF
2.2 nF
RA3
:RANG 2.2NF
4.7 nF :RANG 4.7NF
10 nF :RANG 10NF
100 nF
22 nF
RA4
:RANG 22NF
47 nF :RANG 47NF
100 nF :RANG 100NF
1
µF
220 nF
RA5
:RANG 220NF
470 nF :RANG 470NF
1
µF :RANG 1UF
10
µF
2.2
µF
RA6
:RANG 2.2UF
4.7
µF :RANG 4.7UF
10
µF :RANG 10UF
100
µF
Not
available
RA7 Not available
1 MHz
Auto Auto RB0 :RANG:AUTO ON
1 pF 1 pF RB1 :RANG 1PF
2 pF 2.2 pF RB2 :RANG 2.2PF
4 pF
2.2 pF
RB3
:RANG 2.2PF
4.7 pF :RANG 4.7PF
8 pF
4.7 pF
RB4
:RANG 4.7PF
10 pF :RANG 10PF
16 pF
10 pF
RB5
:RANG 10PF
22 pF :RANG 22PF
32 pF
22 pF
RB6
:RANG 22PF
47 pF :RANG 47PF
64 pF
47 pF
RB7
:RANG 47PF
100 pF :RANG 100PF
128 pF
100 pF
RB8
:RANG 100PF
220 pF :RANG 220PF
256 pF
220 pF
RB9
:RANG 220PF
470 pF :RANG 470PF
512 pF
470 pF
RB10
:RANG 470PF
1 nF :RANG 1NF
1024 pF 1 nF RB11 :RANG 1NF

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4288A and is the answer not in the manual?

Agilent Technologies 4288A Specifications

General IconGeneral
BrandAgilent Technologies
Model4288A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals