Appendix F 269
Error Messages
Error number: -158
F. Error Messages
-158 String data not allowed
A string data element has been received where the 4288A does not accept any string data
element. For example, a parameter must be enclosed with double quotation marks (“...”)
but they are missing.
-160 Block data error
An error not included in the error numbers between -161 and -169 has occurred during the
syntax analysis of block data.
-161 Invalid block data
Block data are expected, but the block data received are invalid for some reason. (Refer to
IEEE488.2,7.7.6.2.) For example, the END message is received before the length of the
block data is reached.
-168 Block data not allowed
A block data element has been received where the 4288A does not accept any block data
element.
-170 Expression error
An error not included in the error numbers between -171 and -179 has occurred during the
syntax analysis of equation data.
-171 Invalid expression
The equation data element is invalid. (Refer to IEEE488.2,7.7.7.2.) For example,
parentheses are not paired or a character violates the standard.
-178 Expression data not allowed
An equation data element has been received where the 4288A does not accept any equation
data element.
-200 Execution error
A comprehensive execution error has occurred for which the 4288A cannot detect further
details. This error code simply indicates the occurrence of an execution error that is defined
in IEEE488.2,11.5.1.1.5.
-211 Trigger ignored
A trigger command or trigger signal has been received and recognized by the 4288A, but it
is ignored due to the timing relationship with the 4288A. For example, this happens when
the 4288A’s trigger system is not in the Waiting for Trigger state.
-213 Init ignored
Another measurement has been being executed and the measurement start request (:INIT
command on page 183) has been ignored.
-214 Trigger deadlock
Indicates that the :READ? command on page 186 was ignored because the trigger source
setting was MAN or BUS.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4288A and is the answer not in the manual?

Agilent Technologies 4288A Specifications

General IconGeneral
BrandAgilent Technologies
Model4288A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals