90 Chapter 7
Sorting Based on Measured Result (Comparator Function)
Sample Program
Lines 530 to 550 Sets the ON/OFF state of the AUX BIN function, the ON/OFF state of
the Low C reject function, and the limit of the Low C reject function to
Aux$, Low_c_rej$, and Low_c_rej_lim.
Lines 560 to 570 Turns ON the comparator function and sets the instrument to display
the comparator sorting result in the instrument setup display area.
Line 610 Sets the trigger mode to the GPIB trigger.
Lines 650 to 680 Turns ON the BIN count function and clears the count values.
Lines 730 to 760 Sets the result display format to the Img$ variable and displays the title
section of the result display.
Lines 770 to 810 Repeats 50 times the operation of triggering the instrument, reading
out the measured result, and displaying the result according to the
Img$ format.
Lines 820 to 850 Reads out the BIN count values.
Lines 860 to 950 Displays the read out BIN count values.