Troubleshooting
6850
6850
Released: April 2004 6850 Series Gas Chromatograph Service Procedures 328
The specifications for the DC power supplies are
noted in Table 1-3.
Table 1-3. Power supply specifications
*Unregulated supply - will fluctuate with AC
power line.
Blown fuses
The following is a list of possible causes for
blown fuses.
AC board:
F3 and F4 protect the GC oven circuit. Possible
cause:
Shorted oven heater element at the hot
side
F1 and F2 protect the primary winding of the
main power transformer. Possible causes:
Shorted primary winding. In this case,
replace the transformer.
A wire from the main power switch or
primary transformer winding shorted to
ground. Could be due to cut insulation or
wire smashed under a screw (usually
after a repair).
Supply % tol Rated I (Fl) System PARD V pp
+24* ±10% 4.75 A < 1 V
–24* ±10% 0.75 A < 1 V
+15 ±5% 0.40 A < 250 mV
–15 ±5% 0.30 A < 250 mV
5 ±2% 4.5 A < 150 mV
40* ±10% 11.25 A (1)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6850 Series and is the answer not in the manual?

Agilent Technologies 6850 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model6850 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals