Lesson 1 – Analog Clock Domain Description
368
Also the clock frequency is set in Clock Domain Setup page. This 
is described in the next lesson.
The selected clock is supplied to all analog modules you set to be 
connected to the analog clock domain. This is independent of the 
card cages in which the modules are installed.
Analog modules in the digital clock domain, independent of the 
card cage they are in, receive their clock from the clock source of 
the digital domain.
Characteristics of Analog Clock Sources and Clock 
Distribution to Analog Modules in Card Cage Slots
INT - The Internal Clock Board of the Analog Card Cage
When you select this analog clock source, the clock board of each 
analog card cage generates the analog domain clock signal for the 
analog modules in its cage. You define the clock signal frequency 
for the analog clock domain in the Clock Domain Setup page.
All clock boards are identical. The clock generation is done by a 64 
bit PLL on the clock boards. This provides a clock resolution of 15 
decimal digits. The clock frequency range is 200 - 500 MHz.
In addition to the generated clock signal, the clock boards can also 
forward the digital domain clock signal to all analog boards at the 
same time. This is summarized in the figure on page 369.
AMC - The Alternate Master Clock 
Mixed-signal tests may require a clock source with lower noise and 
a higher precision than provided by the internal clock board PLL. 
The Alternate Master Clock option (AMC) provides a very stable 
clock signal with extremely low noise and jitter. This signal can be 
directly connected to all analog card cages. The AMC signal is fed 
into the clock board of the card cage. The clock board distributes 
the signal to all slots of the card cage. This is shown in the figure 
below.