Lesson 7 – Executing the DAC Distortion Test
274
In this Lesson ...
Overview You will now learn how to set up and execute the DAC Distortion
test. For the DAC Distortion test, you will apply a sine wave signal,
capture the response, and finally calculate the total harmonic
distortion (THD) and the signal-to-noise ratio (SNR).
The Mixed-Signal Tool will be used for displaying the calculated
spectrum.
This lesson covers the following topics:
• Explanation of the standard DAC distortion tests
• Description of the DAC Distortion testfunction parameters
• Uploading and displaying the calculated spectrum with the Mixed-
Signal Tool.
What you will learn After this lesson you will know how set up the DAC Distortion
testfunction, and how to retrieve the test results.
Related documents More details about the testfunction can be found in the manual
• Agilent 93000 SOC Series “Standard Test Function Reference”.
The Mixed-Signal Tool is described in
• Agilent 93000 SOC Series “Test Setup”