CAPINTEC, INC. CAPRAC
®
-t
April 2021 WELL COUNTER INITIALIZATION 6-31
WIPE TESTS SETUP
Before performing any Wipe measurements, Wipe parameters and Locations must be set up.
Wipe Tests Setup provides the following capabilities:
Set the Trigger Levels for Background, Work Areas, Unrestricted Areas, Sealed
Sources and Packages,
Set the Count Times for Background, Work Areas, Unrestricted Areas, Sealed
Sources and Packages,
Select Nuclides to look for in Work Areas, Unrestricted Areas, Sealed Sources and
Packages, and
Add, edit or delete Wipe Locations and location data.
Wipe Tests Setup is accessed in one of two ways:
1 From Figure 6-1 Main Screen:
a Touch the SETUP button. Figure 6-25 Setup Screen will appear.
Figure 6-25 Setup Screen
b Touch the ADVANCED DETECTOR button. The numeric keypad screen will
appear. Input the password (last 3 digits of the Readout serial number) by
touching the appropriate numbers on the keypad.